GSL1600-80X tubing furnace is a precision bench-top furnace using MoSi2 as heating elements. It is widely used for materials or chemical lab to sinter all types of new materials samples under vacuum or other gases condition. 1/4" in diameter vacuum sealing assembly will connect with 1/4" rubber hose . The temperature of GSL1600-80X tubing furnace is controlled by high precision SCR power controller with accuracy +/-1 oC and 51 segments programmable up to 1600oC (2912 oF). Specifications:
Furnace structure Double layer steel casing with air cooling. Two cooling fan inside to keep case temperature oC
Tube Materials: 99.8% High purity Al2O3 Ceramic
Tube Size: ID: 72mm; OD: 80 mm in diameter x 1000 mm length ---- Standard
Refractory lining 1700 grade fiber alumina caste din vacuum
Heating element 1700 grade MoSi2
Max. Working Temperature: 1600 oC (2912 oF)
Normal Working Temperature: 800 -1500 oC (2732 oF)
Temperature Controller:
PID automatic control via SCR ( Silicon Controlled Rectifier) power control, e.g. current limiting phase angle fired the resistor with 51 steps programmable
Please click here to learn why SCR is more accurate than SSC ( Solid state relay)
Max. Heating / Cooling Rate:
5 oC / min above 1200 oC and 10 oC/min below 1200oC
Heating Zone 457 mm ( 18")
Constant Temperature Zone (+/- 1 oC at 1600 oC ) 150 mm (6")
Temperature Accuracy: +/- 1 oC
Input: Single Phase, 220V AC, 50/60 Hz
Output: Max. 5KW
Dimension: 550 L x 450 W x 670 H mm ( 21.7" L x 17.7" W x 26.4" H )
Net Weight: 150 kg ( 330 lbs )
Dimension of shipment: 47" x 45" x 45"(1200L x 1150W x 1150Hmm)
Weight of shipment:
380 Lb (175kg)
Warranty
One year limited warranty for electronic and mechanical part, excluding heating element and ceramic tube damaged by user
Warning
Clickhere to see typical temperature calibration result and general advise click here to learn How to use and maintain Alumina tube in MTI high temperature tube furnace In order obtain accurate temperature inside tube furnace, you shall calibrate temperature before use, please click here to learn how MoSi2 heating element is used in the high temperature under oxidizing atmosphere. It will form the SiO2 film which can keep the element from being melted. During the oxidizing process, the SiO2 protecting film is formed again when the element continues to be used. The MoSi2 heating element must not be used in the temperature between 400 oC and 700 oC for a long time, otherwise the element will be cremated under the strong oxidizing function in the low temperature. How to set up ceramic tube and vacuum flange for MTI Tube furnace
Standard Package with the FurnaceStandard Package with the Furnace
Part No Description Quantity Click pic to order extra
GSL68-01 MoSi2 Heating Element installed installed
GSL68-02 Vacuum Gauge and Connecting Flanges with 1/4" hose fitting. 4 pcs of High Temperature Vacuum Gaskets are included 1 sets
GSL68-03 Pt-Rh to Pt-Rh Thermal Couples ( B Type ) with alumina protective tube
1 set
GSL68-04 Al2O3 Furnace Tube , ( made in USA )
1 pcs
GSL68-05 Foam Al2O3 Tube Block
4 pcs
GSL68-06 Stainless Steel Tube Protecting Cover 1 pair
GSL68-09 Protective Glove 1 pair Viewproduct manual Download catalog flyer About shipping / handling Quick test instruction
|
|
|
|
|
|
|
|