GSL1600-100X tubing furnace is a precision bench-top furnace employing 1800 grade MoSi2 heating elements. It is widely used for materials or chemical lab to sinter all types of new materials samples under vacuum or other gases condition. 1/4" diameter connector of vacuum sealing assembly will connect with 1/4" rubber hose . The temperature of GSL1600-100X tubing furnace is controlled by high precision controller with accuracy +/-1 oC and 30 segments programmable up to 1550oC Specifications:
Furnace Structure Double layer steel casing with air cooling. Two cooling fan inside to keep temperature of outside case oC
Tube Materials: 99.8% high purity Al2O3 ceramic
Tube Size: 4" OD x 3.75"ID x 40" Length
Refractory Lining 1800 oC grade Fiber alumina
Heating elements 1700 grade MoSi2 ( 8 pcs )
Max. Working Temperature: 1600 oC
Normal Working Temperature: 800 -1550 oC
Temperature Controller:
PID automatic control via current limiting phase angle fired the resistor with 30 segments programmable
Max. Heating / Cooling Rate:
5 oC / min above 1200 oC and 5 oC/min below 1200oC
Heating Zone 455 mm ( 18")
Constant Temperature Zone (Hot Zone +/-1 oC at 1700oC) 150 mm
Temperature Accuracy: +/- 1 oC
Vacuum seal kit one pair of stainless steel vacuum seal flange with gauge and valve included 1/4" hose fitting
Input: Single Phase, 220V AC, 50/60 Hz
Output: Max. 9 KW
Dimension: 550 L x 500 W x 700 H mm ( 21.7" L x 19.7" W x 27.5" H )
Net Weight: 160 kg ( 352 lbs )
Dimension of shipment: 47" x 45" x 45"(1200L x 1150W x 1150Hmm)
Weight of shipment:
400 Lb (175kg)
Warranty
One year limited warranty for electronic and mechanical part, excluding heating element and ceramic tube damaged by user
Warning:
How to use and maintain Alumina tube in MTI high temperature tube furnace In order obtain accurate temperature inside tube furnace, you shall calibrate temperature before use, please click here to learn how MoSi2 heating element is used in the high temperature under oxidizing atmosphere. It will form the SiO2 film which can keep the element from being melted. During the oxidizing process, the SiO2 protecting film is formed again when the element continues to be used. The MoSi2 heating element must not be used in the temperature between 400 oC and 700 oC for a long time, otherwise the element will be cremated under the strong oxidizing function in the low temperature. How to set up ceramic tube and vacuum flange for MTI Tube furnace
Standard Package with the Furnace
Part No Description Quantity Picture
GSL7100-01 1800 grade MoSi2 heating elements 8 pcs in furnace 8 pcs
GSL7100-02 Vacuum Gauge and Connecting Flanges with 1/4" hose fitting. 4 pcs of High Temperature Vacuum Gaskets included 1 sets
GSL7100-03 Pt-Rh to Pt-Rh Thermal Couples ( B Type ) with alumina protective tube
1 set
GSL7100-04 Al2O3 Furnace Tube
1 pcs
GSL7100-05 Foam Al2O3 Tube Block
4 pcs
GSL7100-06 Stainless Steel Tube Protecting Cover installed 1 pair
GSL7100-09 Protective Glove 1 pair View product manual Download catalog flyer About shipping / handling Quick test instruction
|
|
|
|
|
|
|
|